חיישן אמוניה NH3 יציאה 4-20mA
Analog parameter meaning
Current signal output conversion calculation
For example, the range is 0~100ppm, and the output is 4~20mA.
When the output signal is 12mA, calculate the current ammonia concentration.
The span of this ammonia range is 100ppm, which is expressed by a 16mA current signal, 100ppm/16mA=6.25ppm/mA, that is, a current of 1mA represents an ammonia concentration change of 6.25ppm, and the measured value is 12mA 4mA=8mA, 8 mA*6.25ppm/mA =50ppm, the current ammonia concentration is 50ppm.